Safety
Meets the requirements of IEC60601 series, CE marking according to MDD93/42/EC
Size and weight
Dimensions: 310 x 280 x 150mm
Weight: 5.5Kg not including external multi-gas module
Operation environment
Power: AC100~250V, 50/60Hz
Temperature: 0~40C
Humidity: 15~95%
Performance Specifications
Display: 10.4" TFT Display, 800x600 pixels
Trace: 8 waveforms
Sweep speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
Indicator: Alarm indicator, Power indicator, QRS beep and alarm sound
Networking: LAN/WLAN to Mindray Hypervisor III central system
WLAN: Operates on license free ISM (Industrial, Scientific and medical) 2.4 ~ 2.43Ghz Radio frequency 802.11b compliant
Interface: Net port
Battery: Rechargeable; Lead-Acid >45 minutes; Li-Ion >90 minutes
Trend time: 1~96 hours
Alarm: 10 level audible and visual alarm
Recorder: Built-in, thermal array, 2 channels; Record width 48mm (record paper 50mm); record speed 25mm/s, 50mm/s
ECG
5 lead or 3 lead selectable.
Lead mode:
3 lead; RA;LA;LL or R;L;F
5 lead; RA;LA;RL;LL;V or R;L;N;F;C
Lead selection:
3 lead; I, II, III
5 lead; I, II, III, avR, avL, avF, Vx I, II, III, aVR, aVL, aVF, V(n)
ECG waveform: 2 channels
Gain selection: x0.25; x0.5; x1; x2(mm/mv); auto
Sweep speed: 12.5mm/s; 25mm/s; 50mm/s
Heart rate range: 15~350bpm
Accuracy: +/- 1bpm or 1% whichever is the greater
Resolution: 1bpm
Protection: Withstand 4000VAC/50hz voltage in isolation against Electrosurgical interferance and defibrillation
CMRR: Diagnostic mode >90db; Monitor mode >105db; Surgery mode >105db
Operation modes: Diagnostic, monitor, surgery
Bandwidth: 0.05~100hz
Pace detection: YES
Alarm range: 15~350bpm
S-T detection: YES; Measurement range -2.0mV~2.0mV; Alarm range -2.0mV~2.0mV
Arrythmia analysis: YES
Alarm: YES; audible and visual alarm; alarm events recallable
Respiration
Method: RA-LL impedence
Measurement range: 6~150rpm
Accuracy: +/-2rpm
Alarm range: 0~150rpm
Apnea alarm: 10~40s
NIBP
Method: Oscillometric
Operation Mode: Manual/Automatic/STAT
Measurement unit: mmHg/Kpa selectable
Measurement types: Systolic, Diastolic, Mean
Measurement range: 10~270mmHg
Over-pressure protection: YES
Resolution: 1mmHg
Accuracy: Mean error and standard deviation per ANSI/AAMI SP-10
Alarm: Systolic, Diastolic, Mean
Temperature
Measurement range: 0~50C
Resolution: 0.1C
Accuracy: +/-0.1C (not including probe)
Channel; Dual channel
Alarm range: 0~50C
SP02
Measurement range: 0~100%
Resolution: 1%
Accuracy: (70~100%)+/-2%; 0~69% unspecified
Alarm range: 0~100%
Pulse rate
Range: 0~254bpm
Resolution: 1bpm
Accuracy: +/-2bpm or 2% whichever is greater
Alarm range: 0~254bpm