Specialists in Veterinary and Healthcare equipment Telephone :: 0208 763 9777
HOME VETERINARY HEALTHCARE
In this section...
PRODUCTS
SUPPLIES
<<<Back

Safety
Meets the requirements of IEC60601 series, CE marking according to MDD93/42/EC

Size and weight
Dimensions: 310 x 280 x 150mm
Weight:  5.5Kg not including external multi-gas module

Operation environment
Power:  AC100~250V, 50/60Hz
Temperature: 0~40C
Humidity: 15~95%

Performance Specifications
Display: 10.4" TFT Display, 800x600 pixels
Trace:  8 waveforms
Sweep speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
Indicator: Alarm indicator, Power indicator, QRS beep and alarm sound
Networking: LAN/WLAN to Mindray Hypervisor III central system
WLAN:  Operates on license free ISM (Industrial, Scientific and medical) 2.4 ~ 2.43Ghz Radio frequency 802.11b compliant
Interface: Net port
Battery: Rechargeable; Lead-Acid >45 minutes; Li-Ion >90 minutes
Trend time: 1~96 hours
Alarm:  10 level audible and visual alarm
Recorder: Built-in, thermal array, 2 channels; Record width 48mm (record paper 50mm); record speed 25mm/s, 50mm/s


ECG
5 lead or 3 lead selectable.
Lead mode:
3 lead; RA;LA;LL or R;L;F
5 lead; RA;LA;RL;LL;V or R;L;N;F;C
Lead selection:
3 lead; I, II, III
5 lead; I, II, III, avR, avL, avF, Vx I, II, III, aVR, aVL, aVF, V(n)
ECG waveform:  2 channels
Gain selection:  x0.25; x0.5; x1; x2(mm/mv); auto
Sweep speed:  12.5mm/s; 25mm/s; 50mm/s
Heart rate range: 15~350bpm
Accuracy:  +/- 1bpm or 1% whichever is the greater
Resolution:  1bpm
Protection:  Withstand 4000VAC/50hz voltage in isolation against Electrosurgical interferance and defibrillation
CMRR:   Diagnostic mode >90db; Monitor mode >105db; Surgery mode >105db
Operation modes: Diagnostic, monitor, surgery
Bandwidth:  0.05~100hz
Pace detection:  YES
Alarm range:  15~350bpm
S-T detection:  YES; Measurement range -2.0mV~2.0mV; Alarm range -2.0mV~2.0mV
Arrythmia analysis: YES
Alarm:   YES; audible and visual alarm; alarm events recallable

Respiration
Method:   RA-LL impedence
Measurement range: 6~150rpm
Accuracy:  +/-2rpm
Alarm range:  0~150rpm
Apnea alarm:  10~40s

NIBP
Method:    Oscillometric
Operation Mode:   Manual/Automatic/STAT
Measurement unit:  mmHg/Kpa selectable
Measurement types:  Systolic, Diastolic, Mean
Measurement range:  10~270mmHg
Over-pressure protection: YES
Resolution:   1mmHg
Accuracy:   Mean error and standard deviation per ANSI/AAMI SP-10
Alarm:    Systolic, Diastolic, Mean

Temperature
Measurement range: 0~50C
Resolution:  0.1C
Accuracy:  +/-0.1C (not including probe)
Channel;  Dual channel
Alarm range:  0~50C

SP02
Measurement range: 0~100%
Resolution:  1%
Accuracy:  (70~100%)+/-2%; 0~69% unspecified
Alarm range:  0~100%

Pulse rate
Range:  0~254bpm
Resolution: 1bpm
Accuracy: +/-2bpm or 2% whichever is greater
Alarm range: 0~254bpm